T100 Medidor de espesor de revestimiento

Category:

Description

Specification
T100
 

Probe selection

Optional 10 probes
F1(Paint/Plating/Fire-resistant layer and 10mm inner cover)
Other probes: F400,F1/90°,F5,N400,N1,N1/90°,CN02,N10
Working principle
Magnetic induction and eddy current
Measurement accuracy
F1(±((1~2)%H+1)), F10(±((1~3)%H+1)), N1(±((1~2)%H+1))
Measuring range
F1(0~1250um), F10(0~10000um), N1(0~1250um)
Minimum curvature
F1(3mm),F10(10mm),N1(4mm),F400(1mm)
Minimum area
F1(Φ7),F10(Φ40),N1(Φ5),F400(Φ3)
Measurement method
Continuous measurement method, single measurement method
Statistics function
Mean, Large, Small, Tests,Standard Deviation
Calibration method
Single-point calibration,two-point calibration,and five-point basic calibration method to correct the system error of the probe
Storage function
Can store 495 measured values,with delete function:delete single suspicious data appearing in the measurement, or delete all data in the storage area for new measurement
Alam system
Linits can be set: automatic alarming for measured values outside the limits; with power undervoltage indication function; beepprompts during operation; with error prompt function, error prompts can be displayed on screen or by beeps
Shutdown method
Manual shutdown mode and automatic shutdown mode
Use environment
0~40℃/ humidity: 20%RH~90%RH

Magnetic field-free environment
Power supply
Tow AA batteries
The standard configuration
Main unit 1 set/probe 1 set/ stan dard piece(5 pirces)/iron or aluminum base 1 piece/ 5 AA battery 2 sections/manual certificate
warranty card 1 set